Narishige Pc10 Manual New ((better)) -
Ensure the power outlet is properly grounded and matches the voltage specified on the device’s rear serial plate (typically 100V, 120V, or 220-240V).
Insert the filament leads into the terminal blocks, centering the loop perfectly around the vertical glass trajectory path.
What is your specific (e.g., patch-clamp, microinjection, IVF)? Share public link
Secure the glass capillary tubes using knurled thumb screws. The lower slider moves vertically via gravity. narishige pc10 manual new
Center the loop precisely around the central vertical axis where the glass capillary travels.
If you need to program specific pipette dimensions, let me know the or your target tip resistance (MΩ) , and I can suggest baseline heater and weight settings. Share public link
Follow these protocol steps to execute a standard two-stage pull. Step 1: Preparation and Glass Loading Ensure the power outlet is properly grounded and
to the slider shafts; lubricants attract dust and create drag, ruining pull consistency. If necessary, use a dry Teflon-based lubricant. Replacing the Platinum Heating Coil
The is an indispensable tool for electrophysiologists. By understanding the interaction between heater settings and magnet forces, users can generate high-quality electrodes consistently. Always refer to the specific Narishige PC-10 manual that accompanied your unit for detailed safety warnings and precise electrical requirements.
Clean the guide rails with a lint-free cloth and apply a microscopic drop of high-grade machine oil. Preventive Maintenance Share public link Secure the glass capillary tubes
Select your desired pulling mode using the (Set to 2 for a two-stage pull).
Turn on the main power switch. The indicator light should illuminate.
Install the platinum heater coil (Pt-Ir wire). Handle with care, as it is fragile.